High performance differential Dead weight tester BDH3-1200B

huile
balance BDA3 detail 02 balance BDA3 detail 03

reference : BDH3-1200B

Technical specifications

  • measuring : 2 to 1200 bar / 30 to 16000 PSI / 200 to 120000 kPa
  • uncertainly of the pressure measured by the dead weight : 0.002 + (2.10-4 x P* ) (*with P in bar)
  • accuracy : 0.02% of the scale / option : 0.007% if calibration COFRAC or DKD
  • weight of the DWT without weights : 40 Kg
  • fluid : Oil H40 - fluid density : 860 kg/m3
  • typical piston section : 3.92450mm²
  • piston/cylinder/pipe material : P = carbide tungstene / C = treated steel

Options

To be specified when ordering :
- Standard gravity (9.80665m/s²) or special gravity (indicate the desired value)

Options available :
- Intermediaite set of weights 
- A calibration certificate COFRAC or DKD/DAkkS.


additional informations  :

  • Delivery details : a manual are provided in English and French with the device + a factory calibration certificate traceable to the national standards + the data sheet + 0.5 liter of mineral Oil with safety sheet
  • Maintenance : Our technical service department is at your disposition for the revision, calibration or service after sale of your unit – Calibration device recommended every 2 years depending on your use.

other products

Cleaning bench OP0062

Cleaning bench OP0062

  • reference : OP0062

Cleaning bench for pressure gauges

See the details
Standards o-rings for sealing ring

Standards o-rings for sealing ring

  • reference : OP0038CEO

Standards o-rings for sealing ring

See the details
1 liter oil H40
huile

1 liter oil H40

  • reference : OP0014

1 liter oil H40

See the details