Dead weight tester high performance BH2-700B DP

huile
balance BH2 1200B balance BH2 1200B detail 02 balance BH2 1200B detail 01

reference : BH2-700B DP

Technical specifications

  • measuring : 1 to 700 bar / 10 to 10000 PSI / 100 to 70000 kPa
  • uncertainly of the pressure measured by the dead weight : Piston BP (1-70 bar) = 0.001 + 1x10-4 x P* / Piston HP (60-700 bar) = 0.002+1x10-4 x P* (*with P in bar)
  • accuracy : 0.01% of the scale / option : 0.007% if certificate COFRAC or DKD
  • weight of the DWT without weights : 22 Kg
  • generator base : CH2-1250B
  • fluid : Oil H40 - fluid density : 860 kg/m3
  • typical piston section : BP : 78.4472mm² / HP : 7.84068mm²
  • piston/cylinder/pipe material : P:BP = treated steel + HP = carbide tungstene / C = treated steel
  • number of pistons : 2

Options

To be specified when ordering :
- Choice of the unit of the principal set of weights: Bar, PSI, KPa etc...
- Standard gravity (9.80665m/s²) or special gravity (indicate the desired value)

Options available :
- Intermediaite set of weights and/or additional principael set of weights for different units
- A calibration certificate COFRAC or DKD/DAkkS.
- Option OP0004 : Motorization (drive of rotating masses) / Visualization (piston position by indicator lights)
- Option OP0301 : Calculator module with dynamic visualization of the position of the piston, automatic display of the pressures generated by the dead weight according to the temperature, the atmospheric pressure and the hygrometry, with data transfer to a computer for total traceability and calibration certificate edition.
- Adaptation of corrosive products such as skydroll, brake fluid, etc.


additional informations  :

  • Delivery details : a manual are provided in English and French with the device + a factory calibration certificate traceable to the national standards + the data sheet + 0.5 liter of mineral Oil with safety sheet
  • Maintenance : Our technical service department is at your disposition for the revision, calibration or service after sale of your unit – Calibration device recommended every 2 or 5 years depending on your use.

other products

O-rings kit for BH2-300B DP / BH2-600B DP / BH2-1200B DP
huile

O-rings kit for BH2-300B DP / BH2-600B DP / BH2-1200B DP

  • reference : OP0136

O-rings kit for BH2-300B DP / BH2-600B DP / BH2-1200B DP

See the details
1 liter oil H40
huile

1 liter oil H40

  • reference : OP0014

1 liter oil H40

See the details
Separator OP003

Separator OP003

  • measuring : from 0 bar to 800 bar
  • reference : OP0023

Separator liquid/liquid 800 bar

See the details