Dead weight tester high performance BH2-1200B DP

huile
balance BH2 1200B balance BH2 1200B detail 02 balance BH2 1200B detail 01

reference : BH2-1200B DP

Technical specifications

  • measuring : 1 to 1200 bar / 10 to 16000 PSI / 100 to 120000 kPa
  • uncertainly of the pressure measured by the dead weight : Piston BP (1-60 bar) = 0.001 + 1x10-4 x P* / Piston HP (60-1200 bar) = 0.003+1x10-4 x P* (*with P in bar)
  • accuracy : 0.02% of the scale / option : 0.007% if certificate COFRAC or DKD
  • weight of the DWT without weights : 22 Kg
  • generator base : CH2-1250B
  • fluid : Oil H40 - fluid density : 860 kg/m3
  • typical piston section : BP : 78.4472mm² / HP : 3.91903mm²
  • piston/cylinder/pipe material : P:BP = treated steel + HP = carbide tungstene / C = treated steel
  • number of pistons : 2

Options

To be specified when ordering :
- Choice of the unit of the principal set of weights: Bar, PSI, KPa etc...
- Standard gravity (9.80665m/s²) or special gravity (indicate the desired value)

Options available :
- Intermediaite set of weights and/or additional principael set of weights for different units
- A calibration certificate COFRAC or DKD/DAkkS.
- Option OP0004 : Motorization (drive of rotating masses) / Visualization (piston position by indicator lights)
- Option OP0301 : Calculator module with dynamic visualization of the position of the piston, automatic display of the pressures generated by the dead weight according to the temperature, the atmospheric pressure and the hygrometry, with data transfer to a computer for total traceability and calibration certificate edition.
- Adaptation of corrosive products such as skydroll, brake fluid, etc.


additional informations  :

  • Delivery details : a manual are provided in English and French with the device + a factory calibration certificate traceable to the national standards + the data sheet + 0.5 liter of mineral Oil with safety sheet
  • Maintenance : Our technical service department is at your disposition for the revision, calibration or service after sale of your unit – Calibration device recommended every 2 or 5 years depending on your use.

other products

Suitcase with 17 standards adaptors

Suitcase with 17 standards adaptors

  • measuring : from 0 bar to 1200 bar
  • reference : OP0037

Suitcase with 17 standard adaptors 1200 bar

See the details
Option moto-visu OP0004

Option moto-visu OP0004

  • reference : OP0004

Option motorization and visualization with Leds for your dead weight tester range BH2 or BA2

See the details
Micro adjustment valve

Micro adjustment valve

  • reference : OP0194

Micro adjustment valve

See the details